User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Micronta 63-892:Programmable Timer DATE: 09-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 58 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 5335A STANDARD: Stanford Research FS700 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD Stanford Research FS700 1.004 HEAD * INITIAL CONDITIONS * 1.005 DISP The following test accessories are required: 1.005 DISP Timer Calibration Fixture (cabinet 11, C-3) 1.006 DISP Plug the UUT into AC power. 1.007 DISP Set the UUT front panel controls as follows: 1.007 DISP RUN/PROG/TIME SET switch .......................... PROG 1.007 DISP ON/AUTO/OFF switch ................................ AUTO 1.008 DISP Plug TIMER CALIBRATION FIXTURE AC plug into the UUT 1.008 DISP AC receptacle. 1.009 DISP Set the TIMER CALIBRATION FIXTURE controls as follows: 1.009 DISP OPEN/OFF - SHORT/ON switch .................... SHORT/ON 1.010 DISP Connect the Timer Calibration Fixture as follows: 1.010 DISP VOLTAGE IN ....................... 5700A V/Ohm/A OUTPUT 1.010 DISP TO COUNTER ....................... 5335A INPUT A 1.011 HEAD {* TIMER ACCURACY TEST *} 1.012 JMP 2.001 1.013 EVAL 2.001 DISP Use the UUT DAY, HOUR, MIN controls as necessary to set 2.001 DISP a 10 minute on/off program (set ON time then, press 2.001 DISP EVENT and set OFF time for 10 minute later than ON time) 2.002 5700 5V S 2W 2.003 IEEE [@5335][TERM LF] 2.004 IEEE [@5335]IN[D2000] 2.005 IEEE [@5335]FN2, AF1, CO1, TR1, AS1, AT1.25, BT4.75; 2.006 DISP The following instructions must be performed at least 2.006 DISP 30 seconds before programmed ON time begins. 2.006 DISP 2.006 DISP Set the UUT front panel controls as follows: 2.006 DISP RUN/PROG/TIME SET switch ........................... RUN 2.007 HEAD [27][91]7m[27][91]5mPlease Wait, test in progress. 2.008 IEEE [@5335][T0][I] 2.009 ACC 600.0000T 1P% 2.010 HEAD 2.011 MEME 2.012 MEMC T 0.1000% 3.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.